Proline Promag 50W, 53W
Endress+Hauser 13
Performance characteristics
Reference operating
conditions
To DIN EN 29104 and VDI/VDE 2641:
Medium temperature: +28 °C ± 2 K
Ambient temperature: +22 °C ± 2 K
Warm-up period: 30 minutes
Installation:
Inlet run >10 x DN
Outlet run > 5 x DN
Sensor and transmitter grounded.
Sensor centred relative to the pipe.
Maximum measured error Promag 50:
Pulse output: ± 0.5% o.r. ± 1 mm/s (o.r. = of reading)
Current output: plus typically ± 5 µA
Promag 53:
Pulse output: ± 0.2% o.r. ± 2 mm/s (o.r. = of reading)
Current output: plus typically ± 5 µA
Supply voltage fluctuations have no effect within the specified range.
F06-5xxxxxxx-05-xx-xx-xx-000
Max. measured error in % of reading
Repeatability max. ± 0.1% o.r. ± 0.5 mm/s (o.r. = of reading)
0
1
2.5
[%]
2.0
1.5
1.0
0.5
0
246810
v [m/s]
0.2 %
0.5 %